The Thermco Dry Ice Multi-Use Ultra-Low Temperature Data Logger is a multi-trip device that records temperature readings simultaneously. Intended for use in transit monitoring of articles stored in packaging incorporating dry ice cooling agents.
Features:
This device is "Probe-less"
Rated for up to a total cumulative exposure of 1,000 hours at -80°C
‘OK’ & ‘ALERT’ Indicator with high and low customizable alarms
Red Light (ALERT) - Provides an immediate indication, without access to a PC, if any readings are outside the limits specified at the time the unit was configured
Green Light (OK) - Provides immediate visual confirmation, without access to a PC, that the unit is operating
Made with a robust and durable polycarbonate case
Using the LogTag interface and free LogTag analyzer software, you can easily configure start delay, sampling rate, reading count, and ALERT settings
Readings are downloaded with LogTag Analyzer, offering charting, zooming, data statistics, and export options (e.g., Excel)
Technical Specifications:
Temperature Measurement Range: -80°C to 40°C
Operating Temperature Range: -80°C to 40°C
Storage Temperature Range: -20°C to 40°C
Rated Temperature Reading Accuracy: Actual performance is typically much better than the rated values. Accuracy figures can be improved by recalibration
Better than ±1°C for -30°C to 20°C
Better than ±1.2°C for -45°C to -30°C and 20°C to 40°C
Better than ±1.7°C for -80°C to -45°C
Rated Temperature Reading Resolution: LogTag Analyzer currently displays to one decimal place of °C. The native resolution is what is stored in the LogTag®
Less than 0.1°C for -80°C to 0°C
Less than 0.2°C for 0°C to 20°C
Less than 0.5°C for 20°C to 40°C
Memory: 8,003 temperature readings. 53 days at 10-minute logging, 80 days at 15-minute logging
Sensor Reaction Time: Typically less than 5 minutes (T90) in moving air (1m/s)
Sampling Interval: Configurable from 1 minute to several hours
Logging Start Options: Push button start or specific date and time
Download Time: Typically less than 5 seconds for full memory, depending on the computer or readout device used
Power Source: 3.6V low-temperature chemistry Lithium battery (Fixed)
Battery Life: Up to 1 year of storage life before ‘start’ (If it is in hibernation mode)
Real Time Clock:
Rated accuracy ±25 ppm at 25°C (equivalent to 2.5 seconds/day)
Rated temperature coefficient is -0.034 ±0.006 ppm/°C (I.e typically +/- 0.00294 seconds/day/°C)
Certifications: FDA 21 CFR Part 11, EC Annex 11, IATA’s 2018 Dangerous Goods Regulations 59th Edition, FCC, CE, RCM, and RoHS
Warranty: Thermco Products Inc ("The Company", "Us") warrants that your Data Logger, when purchased from an authorized Thermco Products Inc distributor, will be free from defects in materials and workmanship for a period of one (1) year ("Warranty Period") on delivery from our warehouse. This limited warranty is applicable only to the original purchaser ("Customer") who purchased the product from an authorized distributor and is non-transferable.