The Thermco Dual-Channel Ultra-Low Temperature Data Logger with LCD Display is ideal for monitoring and recording temperatures of vaccines in refrigerators or freezers.
Features:
Large, easy-to-read display to see the current, min, and max temperatures
Audible alarm and flashing red LED signal a temperature alarm
View alarm status for up to 30 days at a glance
Real-time data automatically uploaded via WiFi to the LogTag Online cloud
Download data via LogTag Analyzer or as a PDF (no software required) when WiFi is unavailable
Interchangeable external probe with a high-quality gold-plated connector with a stainless steel air probe
Record inspections at the push of a button
Auto-pause function suspends alarm and statistic processing when the external sensor is unplugged
Second battery backup via a coin cell to ensure no data is lost even if the power source and alkaline batteries fail
Comprehensive customization options, including sampling interval and alarm settings
Made with a robust and durable polycarbonate case
Meets and exceeds CDC and VFC with NIST Traceable Certification
Comes with 3m probe and a buffer bottle
Technical Specifications:
Temperature Measurement Range: -90°C to 40°C
Operating Temperature Range: -30°C to 60°C
Storage Temperature Range: 0°C to 40°C
Rated Temperature Reading Accuracy:
Better than ±0.5°C for -20°C to 10°C
Better than ±0.7°C for -50°C to -20°C and 10°C to 30°C
Better than ±1°C for -80°C to -50°C and 30°C to 40°C
Better than ±1.8°C for -90°C to -80°C
Rated Temperature Reading Resolution: LogTag Analyzer® currently displays to one decimal place of °C. The native resolution is what is stored in the LogTag®
0.1°C for measurements -80°C to 10°C
0.2°C for measurements 10°C to 20°C and -90°C to -80°C
0.3°C for measurements 20°C to 30°C
0.5°C for measurements 30°C to 40°C
Memory: 16,129 temperature readings. 67 days at 6-minute logging
Sensor Reaction Time: Typically less than 2 minutes(T90) in moving air (1m/s) for ST10S type
Sampling Interval: Configurable from 30 seconds to 18 hours
Logging Start Options: Push button start (with configurable start delay from 1 minute to 72 hours) or specific date and time
Download Time: Typically, with full memory, in less than 10 seconds
Power Source: 5V via micro USB socket or 2 x 1.5V AAA alkaline batteries. CR2032 LiMg coin cell (backup logging function only)
Battery Life: AAA battery up to 12 months, if the unit is operated without a power supply
Real Time Clock:
Rated accuracy ±25 ppm at 25°C (equivalent to 2.5 seconds/day)
Rated temperature coefficient is -0.034 ±0.006 ppm/°C
Warranty: Thermco Products Inc ("The Company", "Us") warrants that your Data Logger, when purchased from an authorized Thermco Products Inc distributor, will be free from defects in materials and workmanship for a period of one (1) year ("Warranty Period") on delivery from our warehouse. This limited warranty is applicable only to the original purchaser ("Customer") who purchased the product from an authorized distributor and is non-transferable.