CT Head with 5 slices to test all parameters of computerized tomography
Anthropomorphic
Molded Around Skull
Checks Important Physical Parameters
Provides Realistic System Check
Ideal For Training
Polycarbonate Base Dimensioned For Scans Parallel or Perpendicular to Longitudinal Axis
SECTION A: Aluminum plates, 0.4mm thick, set at a 45° angle, measure beam width and slice thickness.
SECTION B: The dosimetry section provides patient-exposure controls, and checks on abnormal internal doses before actual patient scans. Customization to accommodate TLD with rods or chips is available.
SECTION C: Cylindrical tumors of graded sizes and radiodensities establish high and low-contrast resolution and demonstrate partial beam-averaging effects.
SECTION D: The anthropomorphic section provides actual auditory ossicles in an air-contrast inner ear and a fractured petrous bone. Small tumors in the posterior fossa are placed in an area which most scanners image with difficulty.
SECTION E: 1/4 Inch diameter aluminum rod on the phantom's longitudinal axis, checks general alignment, and "DryWater" gives a realistic "noise check."
Availability: Varies; Please contact Customer Service if you require a specific delivery date.
Country of Origin: United States of America
About this Brand
After developing the first anthropomorphic test dummy for the aviation industry in the late 1940's, the RSD team directed its impactful R & D to revolutionize healthcare diagnostic imaging with anthropomorphic phantoms. RSD diagnostic imaging phantoms are routinely used to perform quality control on imaging equipment and to train healthcare professionals in best practices. The company has become the leading supplier of QC and educational solutions for medical diagnostics and radiation therapy, featuring products such as anthropomorphic phantoms and dosimetry verification phantoms.