Mindray ePM 15M Patient Monitor with 15.60" Touch Screen
-
Mindray Sidestream CO2 Module for ePM 10M, 12M, and 15M Patient MonitorsMSRP $3,700.00 MSRP $3,700.00 $2,378.00 -
Mindray Oridion® Microstream® CO2 Module for ePM 10M, 12M, and 15M Patient MonitorsMSRP $4,440.00 MSRP $4,440.00 $2,854.00
Product Description
The Mindray ePM 15M Patient Monitor with 15.60" Touch Screen offers a plug-and-play module structure for more advanced monitoring functions such as CO2, aiding clinicians in identifying signs of patient deterioration due to respiratory distress.
Features:
- 12 Standard waveforms
- Integrated parameters include 3-, 5-lead ECG, SpO2, NIBP, IBP, respiration, and temperature, offering a robust parameter platform to meet the needs of multiple care environments
- Arrhythmia analysis with atrial fibrillation (Afib) detection standard
- Extensive data storage capability for trend data, alarms, events, and 48 hours of full disclosure
- Visual and Operational Experience:
- 15.60" Smartphone-like multi-gesture, capacitive touchscreen display
- Flat, intuitive user interface design, common across platforms
- HD display with wide-angle view ≥ 170
- Reach 90% of common monitor functions in 2 steps
- Up to 18 configurable quick keys
- Clinician-Focused Technology for Enhanced Patient Care:
- Parameter list tile with automatic highlight of abnormal readings
- One-touch to drag icon tab right and instantly view mini-trend screen or set baseline vitals at the start of a patient case
- Remote view and alarm watch bed-to-bed allows up to 12 patients to be viewed
- Connects to BeneVision DMS and to Mindray eGateway for integrated HL7 and DIAP communication protocols for interface to EMR
- Ergonomic Design Delivers Exceptional Performance and Reliability:
- Manufactured with industry-leading plastics that stand up to the most stringent medical-grade cleaners
- IP44 rated, resistant to damage often caused by continuous cleaning and usage
- Expanding extension rack supports plug-and-play parameter modules
- The concealed grab-and-go handle eliminates hard-to-clean crevices and eases clinical workflow while on the move
- Flexible mounting options for bedside and transport
Technical Specifications:
- Display:
- Type: Capacitive Touchscreen, Supports Multi-Touch Operation
- Size and Resolution: 15.60", 1366 x 768 Pixels
- Waveforms: Up to 12 Waveforms
- ECG (3, 5 Lead):
- Leads: I, II, III, aVR, aVL, aVF, V
- Sweep Speeds: 6.25 mm/s, 12.50 mm/s, 25 mm/s, 50 mm/s
- Gain Selection: x 0.125, x 0.25, x 0.50, x 1, x 2, x 4, Auto
- Input Signal Range: ±8 mV (p-p)
- Electrode Offset Potential Tolerance: ±500 mV
- Bandwidth:
- Diagnostic Mode: 0.05 to 150 Hz
- Monitor Mode: 0.50 to 40 Hz
- Surgical Mode: 1 to 20 Hz
- ST Mode: 0.05 to 40 Hz
- Defibrillator Protection: Withstands 5000 V (360 J) Voltage in Isolation Against Defibrillation
- Recovery Time: ≤ 5 s
- CMRR:
- Diagnostic: > 90 dB
- Monitor, Surgical, ST Mode: > 105 dB (with Notch Filter On)
- ESU Recovery Time: ≤ 10 s
- Heart Rate Meter:
- Measurement Range:
- Adult: 15 to 300 BPM
- Pediatric/Neonate: 15 to 350 BPM
- Accuracy: ±1 BPM or ±1%, Whichever is Greater
- Resolution: 1 BPM
- Measurement Range:
- Pace Pulse:
- Pace: Pace Pulses Meeting the Following Conditions are Labelled with a PACE Marker:
- Amplitude: ±2 to ±700 mV
- Width: 0.10 to 2 ms
- Rise Time: 10 to 100 μs (Without Overshoot)
- Pace Pulse Rejection: When Tested in Accordance with the ANSI/AAMI/IEC 60601-2-27: 201.12.1.101.13, the Heart Rate Meter Rejects All Pulses Meeting the Following Conditions
- Amplitude: ±2 to ±700 mV
- Width: 0.10 to 2 ms
- Rise Time: 10 to 100 μs (Without Overshoot)
- Pace: Pace Pulses Meeting the Following Conditions are Labelled with a PACE Marker:
- Arrhythmia Analysis
- Patient: Adult/Pediatric
- Monitored Arrhythmias: Asystole, V-Fib/V-Tach, V-Tach, Vent Brady, Extreme Tachy, Extreme Brady, Vent Rhythm, PVCs/min, Pauses/min, Couplet, Bigeminy, Trigeminy, R on T, Run PVCs, PVC, Tachy, Brady, Missed Beats, Pacer Not Pacing, Pacer Not Capture, Multiform PVC, Nonsus V-Tach, Pause, Irr Rhythm, and A-Fib
- ST Segment Analysis (Optional)
- Patient: Adult/Pediatric/Neonate
- Range: -2 to +2 mV (RTI)
- Accuracy:
- -0.80 to +0.80 mV: ±0.02 mV or ±10%, Whichever is Greater
- Beyond This Range: Not Specified
- Resolution: 0.01 mV
- ST Adjust Scale:
- 40 ms After J Point
- 60 ms After J Point (Default)
- 80 ms After J Point
- 60/80 ms After J Point (80 ms when HR is 120 BPM or Less; 60 ms when HR is above 120 BPM)
- ISO Adjust Scale:
- 4 to 200 ms Before R-Wave (Default: 80 ms)
- Step: 4 ms
- J Point Adjust Scale:
- 4 to 200 ms After R-Wave (Default: 48 ms)
- Step: 4 ms
- QT Analysis (Optional)
- Patient: Adult/Pediatric/Neonate
- Parameters: QT, QTc, ΔQTc
- QTc Formula: Bazett, Fridericia, Framingham, and Hodges
- Range:
- QT/QTc: 200 to 800 ms
- QT-HR:
- Adult: 15 to 150 BPM
- Pediatric/Neonate: 15 to 180 BPM
- QT Accuracy: ±30 ms
- Resolution:
- QT: 4 ms
- QTc: 1 ms
- Respiration
- Lead: I or II, Auto
- RR Range:
- 0 to 120 RPM (Adult)
- 0 to 150 RPM (Pediatric/Neonate)
- RR Accuracy:
- ±2 RPM (7 to 150 RPM) or ±2%, Whichever is Greater
- 0 to 6 RPM: Not Specified
- RR Resolution: 1 RPM
- Sweep Speed: 3 mm/s, 6.25 mm/s, 12.50 mm/s, 25 mm/s, 50 mm/s
- Apnea Alarm Delay: 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
- Pulse Oximetry:
- With Masimo SET® SpO2:
- Range: 1 to 100%
- Resolution: 1%
- Accuracy:
- ±2% (70 to 100%, Adult/Pediatric, Non-Motion)
- ±3% (70 to 100%, Neonate, Non-Motion)
- ±3% (70 to 100%, Motion)
- 1 to 69% Unspecified
- Pulse Rate Range: 25 to 240 BPM
- Pulse Rate Accuracy:
- ±3 BPM (Non-Motion)
- ±5 BPM (Motion)
- Refresh Rate: ≤1 s
- With Nellcor Pulse Oximetry:
- Range: 0 to 100%
- Resolution: 1%
- Accuracy:
- ±2% (70 to 100%, Adult/Pediatric)
- ±3% (70 to 100%, Neonate)
- Unspecified (0 to 69%)
- Pulse Rate Range: 20 to 300 BPM
- Pulse Rate Accuracy:
- ±3 BPM (20 to 250 BPM)
- Unspecified (251 to 300 BPM)
- Refresh Rate: ≤1 s
- With Masimo SET® SpO2:
- PR:
- Range:
- 25 to 350 BPM (From IBP)
- 30 to 300 BPM (From NIBP)
- Accuracy:
- ±1 BPM or ±1%, Whichever is Greater (From IBP)
- ±3 BPM or ±3%, Whichever is Greater (From NIBP)
- Range:
- Temperature
- Meets the Standards of ISO 80601-2-56
- Method: Thermal Resistance
- Channels: 2
- Range: 0 to 50°C
- Accuracy: ±0.10°C (Without Probe)
- Resolution: 0.10°C
- Non-Invasive Blood Pressure
- Meets the Standards of ISO 80601-2-30
- Method: Oscillometry
- Operating Mode: Manual, Auto, STAT, Sequence
- Maximum Measurement Time:
- Adult/Pediatric: 180 s
- Neonate: 90 s
- Systolic Range:
- Adult: 25 to 290 mmHg
- Pediatric: 25 to 240 mmHg
- Neonate: 25 to 140 mmHg
- Diastolic Range:
- Adult: 10 to 250 mmHg
- Pediatric: 10 to 200 mmHg
- Neonate: 10 to 115 mmHg
- Mean Range:
- Adult: 15 to 260 mmHg
- Pediatric: 15 to 215 mmHg
- Neonate: 15 to 125 mmHg
- Accuracy:
- Maximum Mean Error: ±5 mmHg
- Maximum Standard Deviation: 8 mmHg
- Resolution: 1 mmHg
- Assisting Venous Puncture: Yes
- Invasive Blood Pressure (Optional)
- Meets the Standards of IEC 60601-2-34:2011
- Technique: Direct Invasive Measurement
- Measurement Range: -50 to 300 mmHg
- Resolution: 1 mmHg
- Accuracy: ±2% or ±1 mmHg, whichever is Greater (Excluding Sensor Error)
- Refreshing Rate: ≤ 1 s
- Recovery Time: < 10 s (After Defibrillation)
- Sidestream CO₂ (Optional)
- Measurement Range: etCO2: 0 to 150 mmHg
- CO2 Accuracy:
- 0 to 40 mmHg: ±2 mmHg
- 41 to 76 mmHg: ±5% of Reading
- 77 to 150 mmHg: ±10% of Reading
- Resolution: etCO2: 1 mmHg
- Sample Flow Rate:
- Adult/Pediatric: 120 mL/min
- Neonate: 70 mL/min or 90 mL/min
- Sample Flow Rate Tolerance: ±15 mL/min or ±15%, Whichever is Greater
- Warm-Up Time: 90 s (Maximum), 20 s (Typically)
- Rise Time: etCO2:
- ≤250 ms @ 70 mL/min (Neonate Water Trap)
- ≤250 ms @ 90 mL/min (Neonate Water Trap)
- ≤300 ms @ 120 mL/min (Adult/Pediatric Water Trap)
- Response Time: etCO2:
- ≤5.0 s @ 70 ml/min (Neonate Water Trap)
- ≤4.5 s @ 90 ml/min (Neonate Water Trap)
- ≤5.0 s @ 120 ml/min (Adult/Pediatric Water Trap)
- awRR Range: 0 to 150 RPM
- awRR Accuracy:
- 0 to 60 RPM: ±1 RPM
- 61 to 150 RPM: ±2 RPM
- Apnea Alarm Delay: 10, 15, 20, 25, 30, 35, 40 s
- Oridion Microstream CO2 (Optional)
- Measurement Range: 0 to 99 mmHg
- Resolution: 1 mmHg
- Accuracy:
- 0 to 38 mmHg: ±2 mmHg
- 39 to 99 mmHg: ±5% of Reading (0.08% Increase in Error for Every 1 mmHg if Reading is More than 38 mmHg)
- Sample Flow Rate: 50 -7.50 +15 mL/min
- Start-Up Time: 30 s (Typical), 180 s (Max)
- awRR Range: 0 to 150 RPM
- awRR Accuracy:
- 0 to 70 RPM: ±1 RPM
- 71 to 120 RPM: ±2 RPM
- 121 to 150 RPM: ±3 RPM
- Alarm: 10, 15, 20, 25, 30, 35, 40 s
- Anesthetic Gas Monitoring (Optional External Module)
- Sampling Rate:
- Adult/Pediatric: 200 mL/min
- Neonate: 120 mL/min
- Sampling Rate Tolerance: ±10 mL/min or ±10%, Whichever is Greater
- Delay Time: < 4 s
- Warm-Up Time:
- 45 s to Iso Accuracy Mode
- 10 min to Full Accuracy Mode
- Measurement Range:
- CO2: 0 to 30%
- N2O: 0 to 100%
- Des/Sev/Enf/Iso/Hal: 0 to 30%
- O2: 0 to 100%
- awRR: 2 to 100 RPM
- Resolution
- CO2: 0.10%
- N2O: 1%
- Des/Sev/Enf/Iso/Hal: 0.10%
- O2: 1%
- awRR: 1 RPM
- Rise Time:
- Sampling Flow of 120 mL/min, using the Neonatal DRYLINE II™ Water Trap and a Neonatal 2.50 m Sampling Line:
- CO2/ N2O: ≤ 250 ms
- Iso/Hal/Sev/Des: ≤ 300 ms
- Enf: ≤ 350 ms
- O2: ≤ 600 ms
- Sampling Flow of 200mL/min, using the Adult/Pediatric DRYLINE II™ Water Trap and an Adult 2.50 m Sampling Line:
- CO2/ N2O: ≤ 250 ms
- Iso/Hal/Sev/Des: ≤ 300 ms
- Enf: ≤ 350 ms
- O2: ≤ 500 ms
- Sampling Flow of 120 mL/min, using the Neonatal DRYLINE II™ Water Trap and a Neonatal 2.50 m Sampling Line:
- Sampling Delay Time:
- Sampling Flow of 120 mL/min, using the Neonatal DRYLINE II™ Water Trap and a Neonatal 2.50 m Sampling Line:
- CO2: ≤ 4 s
- N2O: ≤ 4.20 s
- O2: ≤ 4 s
- Enf /Iso/Hal/Sev/Des: ≤ 4.40 s
- Sampling Flow of 200mL/min, using the Adult/Pediatric DRYLINE II™ Water Trap and an Adult 2.50 m Sampling Line:
- CO2: ≤ 4.20 s
- N2O: ≤ 4.30 s
- O2: ≤ 4 s
- Enf/Iso/Hal/Sev/Des: ≤ 4.50 s
- Apnea Alarm Delay: 10, 15, 20, 25, 30, 35, 40 s
- Sampling Flow of 120 mL/min, using the Neonatal DRYLINE II™ Water Trap and a Neonatal 2.50 m Sampling Line:
- Sampling Rate:
- Data Storage and Review
- Trends Data: Up to 120 hours @ 1 min
- Events: Up to 1,000 Events, Including Parameter Alarms, Arrhythmia Events, and Technical Alarms
- NIBP: Up to 1,000 Sets
- Full Disclosure: 48 hours for All Parameter Waveforms
- OxyCRG: 400 OxyCRG Events
- ST Review: Up to 120 hours @ 1 min
- Wi-Fi Communications
- Protocol: IEEE 802.11a/b/g/n
- Modulation Mode: DSSS and OFDM
- Operating Frequency:
- IEEE 802.11b/g/n (2.4G):
- ETSI/FCC/KC: 2.40 to 2.48 GHz
- MIC: 2.40 to 2.50 GHz
- IEEE 802.11a/n (5G):
- ETSI: 5.15 to 5.35 GHz, 5.47 to 5.73 GHz
- FCC: 5.15 to 5.35 GHz, 5.73 to 5.82 GHz
- MIC: 5.15 to 5.35 GHz
- KC: 5.15 to 5.35 GHz, 5.47 to 5.73 GHz, 5.73 to 5.82 GHz
- IEEE 802.11b/g/n (2.4G):
- Channel Spacing: 5 MHz @ 2.4 GHz, 20 MHz @ 5 GHz
- Wireless Baud Rate:
- IEEE 802.11a: 6 to 54 Mbps
- IEEE 802.11b: 1 to 11 Mbps
- IEEE 802.11g: 6 to 54 Mbps
- IEEE 802.11n: 6.5 to 72.20 Mbps
- Output Power:
- < 20 dBm (CE Requirement: Detection Mode: RMS)
- < 30 dBm (FCC requirement: Detection Mode: Peak Power)
- Operating Mode: Infrastructure
- Data Security: WPA-PSK, WPA2-PSK, WPA-Enterprise, WPA2-Enterprise (EAP-FAST, EAP-TLS, EAP- TTLS, PEAP-GTC, PEAP-MSCHAPv2, PEAP-TLS, LEAP)
- Encryption: AES
- Interfacing:
- Main Unit:
- (1) AC Power Connector
- (1) VGA Port
- (2) USB 2.0 Connector
- (1) Analog Output/Nurse Call/Defib Sync/Serial Port
- (1) Integrated Module Rack, For 2 Slots
- Barcode Scanner: Supports 1D and 2D Barcodes
- Thermal Recorder: 3 Traces (Paper: 50 mm W x 20 m L)
- Network Printer Support: Yes
- Main Unit:
- Power Requirements:
- Line Voltage: 100 to 240 VAC (±10%)
- Maximum Current: 2.0 A
- Frequency: 50/60 Hz (±3 Hz)
- Battery: Rechargeable Lithium-Ion, 4500 mAh
- Battery Runtime: ≥4 Hours
- Recharge Time (Power Off): 5 hours to 90%
- Environmental Requirements
- Operating Temperature
- 0 to 40°C (Without AG and CO2 Module)
- 10 to 40°C (With AG and CO2 Module)
- Operating Humidity: 15 to 95%, Non-Condensing
- Operating Barometric Pressure: 427.5 to 805.5 mmHg, Without Anesthesia Gas and CO2 Modules
- Operating Temperature
- Overall Dimensions: 15.51" L x 6.85" W x 12.44" H
- Actual Product Weight: 13.70 lbs (Standard Configuration, Excluding Modules, Recorder, Battery, and Accessories)
- Warranty: 3 Years
Available Options:
| SKU | SpO2 | ST/QT/QTc Monitoring | Dual IBP | WiFi |
| 121-002690-00 | Nellcor® OxiMax® | - | - | - |
| 121-002668-00 | Masimo SET® | - | - | - |
| 121-002703-00 | Nellcor® OxiMax® | ✓ | ✓ | ✓ |
| 121-002713-00 | Nellcor® OxiMax® | ✓ | - | - |
| 121-002712-00 | Masimo SET® | ✓ | - | - |
| 121-002692-00 | Nellcor® OxiMax® | ✓ | ✓ | - |
| 121-002701-00 | Nellcor® OxiMax® | - | - | ✓ |
| 121-002670-00 | Masimo SET® | ✓ | ✓ | - |
| 121-002679-00 | Masimo SET® | - | - | ✓ |
| 121-002681-00 | Masimo SET® | ✓ | ✓ | ✓ |
Please note: Only the ePM 15M is included. All other items shown are for demonstration purposes only and are not included. The pole is not included.
About this Brand
Specifications
| Manufacturer | MDPro |
|---|---|
| Latex Free | Yes |
| Actual Product Weight | 13.70 lbs |
| Overall Height | 12.44" |
| Overall Width | 6.85" |
| Overall Length | 15.51" |
| Country of Manufacture | China |
| Item Ships From | California |
Literature
Returns
Products must be returned in unused, resalable condition with the original packaging. Please contact Customer Service for a valid return authorization within 30 days of receipt of purchase. A minimum restocking fee of 25% of the purchase price will be deducted from the refund. Customers original shipping charges will not be reimbursed. The customer is responsible for return shipping charges and insuring the package. Credit will not be issued if product is received damaged.