Product Description

The Mindray Anesthetic Gas Module for ePM 12M Patient Monitor measures the patient's anesthetic and respiratory gases by connecting to the airway of intubated patients or by collecting the gases with specified accessories. It is intended for adult, pediatric, and neonatal patients.

Features:

  • Can send waves and numeric data for all measured anesthetic gases for display on the monitor, including:
    • CO2, O2, N2O, and AA Waves
    • awRR: Airway Respiratory Rate
    • MAC: Minimum Alveolar Concentration
    • End-tidal (Et) and Fraction of Inspired (Fi) Numerics for CO2, O2, N2O, and AA

Technical Specifications:

  • Meets the Standards of ISO 80601-2-55:2011
  • Technique: Infrared Absorption, Paramagnetic Properties for O2 Monitoring
  • Warm-Up Time:
    • Iso Accuracy Mode: 45 s
    • Full Accuracy Mode: 10 min
  • Sample Flow Rate:
    • Adult/Pediatric: 200 mL/min
    • Neonate: 120 mL/min
    • Accuracy: ±10 mL/min or ±10%, Whichever is Greater
  • Measurement Range:
    • CO2: 0 to 30%
    • O2: 0 to 100%
    • N2O: 0 to 100%
    • Des/Sev/Enf/Iso/Hal: 0 to 30%
    • awRR: 2 to 100 RPM
  • Resolution
    • CO2: 0.10%
    • O2: 1%
    • N2O: 1%
    • Des/Sev/Enf/Iso/Hal: 0.10%
    • awRR: 1 RPM
  • Iso Accuracy: As Full Accuracy Specifications, But Derated As Follows:
    • Add ±0.30% ABS to Accuracy for CO2
    • Add ±8% REL to Accuracy for All Anesthetic Gases
    • N2O Accuracy is ±(8% REL + 2% ABS)
  • awRR Accuracy:
    • 2 to 60 RPM: ±1 RPM
    • >60 RPM: Not Specified
  • Accuracy Drift: Meet the Requirement for Measurement Accuracy Within 6 Hours
  • Apnea Alarm Delay: 10, 15, 20, 25, 30, 35, 40 s
  • Refreshing Rate: ≤ 1 s
  • Rise Time (10% to 90%):
    • Gas Sample Flow Rate of 120 mL/min, using the Neonatal DRYLINE II™ Water Trap and a Neonatal 2.50 m Sampling Line:
      • CO2/ N2O: ≤ 250 ms
      • Iso/Hal/Sev/Des: ≤ 300 ms
      • Enf: ≤ 350 ms
      • O2: ≤ 600 ms
    • Gas Sample Flow Rate of 200mL/min, using the Adult/Pediatric DRYLINE II™ Water Trap and an Adult 2.50 m Sampling Line:
      • CO2/ N2O: ≤ 250 ms
      • Iso/Hal/Sev/Des: ≤ 300 ms
      • Enf: ≤ 350 ms
      • O2: ≤ 500 ms
  • Response Time:
    • Gas Sample Flow Rate of 120 mL/min, using the Neonatal DRYLINE II™ Water Trap and a Neonatal 2.50 m Sampling Line:
      • CO2: ≤ 4 s
      • N2O: ≤ 4.20 s
      • O2: ≤ 4 s
      • Hal/Iso/Sev/Des/Enf: ≤ 4.40 s
    • Gas Sample Flow Rate of 200mL/min, using the Adult/Pediatric DRYLINE II™ Water Trap and an Adult 2.50 m Sampling Line:
      • CO2: ≤ 4.20 s
      • N2O: ≤ 4.30 s
      • O2: ≤ 4 s
      • Hal/Iso/Sev/Des/Enf: ≤ 4.50 s
  • Anesthetic Agent Limit:
    • Primary Anesthetic Agent:
      • In Full Accuracy Mode: 0.15%
    • Secondary Anesthetic Agent:
      • In Full Accuracy Mode:
        • 5% of Primary Agent if Primary Agent is > 10%
        • 0.30% if Primary Agent is ≤ 10%
  • Data Sample Rate: 25 Hz
  • Overall Dimensions: 5.40" L x 4" W x 3.20" H
  • Actual Product Weight: 3.50 lbs
  • Warranty: 90 Days

About this Brand

Specifications

Specifications
Manufacturer MDPro
Latex Free Yes
Actual Product Weight 3.50 lbs
Overall Height 3.20"
Overall Width 4"
Overall Length 5.40"
Country of Manufacture China
Item Ships From California

Literature

pdf
(Size: 5.3 MB)

Mindray ePM 10M and 12M Patient Monitors

Returns

Products must be returned in unused, resalable condition with the original packaging. Please contact Customer Service for a valid return authorization within 30 days of receipt of purchase. A minimum restocking fee of 25% of the purchase price will be deducted from the refund. Customers original shipping charges will not be reimbursed. The customer is responsible for return shipping charges and insuring the package. Credit will not be issued if product is received damaged.

Reviews

Ask a Question

Customer Questions
This product has no questions.
Did you find what you were looking for?